Scientific poster presentation award
Evaluation of Extracorporeal Blood Heat Exchanger Devices
Division of Experimental Surgery, Ranguell Hospital and National Institute for Health and Medical Research, Hotel Dieu Hospital Toulouse, France
* Address correspondence to: Robert Barthelemy, MD, Div. of Experimental Surgery, Rangueil Hospital, Chemin du Vallon, 31054 Toulouse Cedex, France.
An accurate and easy to use test circuit has been developed for the performance evaluation of the heat exchangers used in cardiac surgery during extracorporeal circulation. The water part of the heat exchanger is fixed at a flow rate of 15 Vmin and at a temperature of 40°C. In the blood side, the flow rate is varied from 1 to 6 1/min and the temperature is 30°C. The use of a precise thermometer (precision 0.01 °C) and of a microcomputer permits efficiency with a good reproducibility. The results obtained show a real improvement in the efficiency of the new extracorporeal blood heat exchangers devices.
Key words: Extracorporeal circulation / heat exchanger / efficiency