Scientific poster presentation award: Evaluation of Extracorporeal Blood Heat Exchanger Devices
Division of Experimental Surgery, Ranguell Hospital and National Institute for Health and Medical Research, Hotel Dieu Hospital, Toulouse, France
* Address correspondence to: Robert Barthelemy, MD, Division of Experimental Surgery, Rangueil Hospital, Chemin du Val/on, 31054 Toulouse Cedex, France.
An accurate and easy to use test circuit has been developed, for performance evaluation of heat exchangers used in cardiac surgery during extracorporeal circulation. The water part of the heat exchanger is fixed at a flow rate of 15 1/min and at a temperature of 40°C. For blood, the flow rate is varied from 1 to 6 1/min and the temperature is 30°C. The use of a very precise thermometer (precision 0.01 °C) and of a microcomputer permits one to obtain the efficiency with very good reproducibility. The results obtained show a real improvement in the efficiency of the new extracorporeal blood heat exchanger devices.
Key words: Extracorporeal circulation / heat exchanger / efficiency
© 1990 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.