Open Access

Table 2

NewYork-Presbyterian Morgan Stanley Children’s Hospital vs. ELSO neonatal and pediatric complication data.

Complication detail Unadjusted rate neonatal ELSO unadjusted rate neonatal Unadjusted rate pediatric ELSO unadjusted rate pediatric
Any mechanical complication 33.80% 31.50% 28.10% 26.40%
 Oxygenator failure 3.08% 5.26% 9.18% 5.13%
 Raceway rupture 0.00% 0.05% 0.00% 0.06%
 Other tubing rupture 0.00% 0.09% 0.00% 0.22%
 Pump malfunction 1.54% 0.70% 0.51% 0.67%
 Heat exchanger malfunction 0.00% 0.22% 0.00% 0.18%
 Clots: hemofilter 0.00% 2.68% 0.00% 2.05%
 Clots: circuit component clots 6.15% 12.00% 7.65% 8.00%
 Air in circuit 6.15% 2.74% 2.04% 2.97%
 Cracks in pigtail connectors 0.00% 0.08% 0.00% 0.13%
 Cannula problems 9.59% 10.20% 3.9% 8.59%
 Circuit change 6.15% 12.60% 8.16% 9.31%
 Clots and air emboli 1.54% 1.02% 0.51% 0.45%
 Thrombosis/Clots: circuit component 4.62% 11.10% 7.14% 8.09%
Any hemorrhagic complication 26.20% 19.50% 25.50% 21.20%
 Cannulation site bleeding 10.80% 12.60% 11.70% 12.80%
 Surgical site bleeding 12.30% 7.32% 13.80% 8.08%
 Hemolysis (PFH > 50 mg/dl) 21.50% 15.30% 9.69% 11.00%

PFH = plasma free hemoglobin.