Implosion/Explosion of a Bentley Cardiotomy Reservoir during Autotransfusion with a Cell Saver
Saint Francis Medical Center, Monroe, LA
* Direct communications to: George E. Cimochowski, M.D., Cardiac Surgery Associates of Northeast Louisiana, Inc., 141 DeSiard, Suite 410, Monroe, LA 71201
An implosion/explosion hazard exists in those institutions using a Bentley BCR 3000° cardiotomy reservoir during autotransfusion if the wall vacuum is applied to the reservoir and the sucker tip becomes occluded for fifty or more seconds. A case report and preventive measures are presented.