Can Vacuum Assisted Venous Drainage be Achieved using a Roller Pump in an Emergency? A Pilot Study using Neonatal Circuitry
Address correspondence to: S.L. Hill, BS, RRT, University of Nebraska Medical Center, Omaha, NB 68198-5155. E-mail: slhill@unmc.edu
There has been much advancement in perfusion technology over its 50 years of progression. One of these techniques is vacuum-assisted venous drainage (VAVD). Many perfusionists augment venous drainage using VAVD, typically from a wall vacuum source. This study explores alternates to providing VAVD if the wall vacuum fails. In two porcine laboratories, ∼36 in. of 3/16-in. tubing was connected to a sucker return port and placed into the roller head next to the arterial pump. The vacuum was monitored with a DLP pressure monitoring system (Medtronic). This system was connected to small-bore tubing and attached to a stopcock on top of the reservoir. The vacuum was regulated using another stopcock connected to a non-filtered luer lock port on top of the reservoir or by a segment of 3 × 0.25-in.-diameter tubing attached to the vent port with a c-clamp. Vacuum drainage was achieved, ranging from −18 mmHg to −71 mmHg by manipulating the stopcock or c-clamp. Changes in venous drainage were seen by volume fluctuations in the venous reservoir. The vacuum was adjusted to account for dramatic changes. Augmented venous drainage using a roller pump can be achieved successfully during cardiopulmonary bypass (CPB). This method of active drainage can be used in lieu of wall suction or during times of emergency if wall suction fails.
Key words: augmented venous drainage / vacuum / roller pump / neonatal circuitry
© 2007 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.