Rupture of Extra-Corporeal Circuit Tubing During Cardiopulmonary Bypass
* Department of Cardio-Thoracic Surgery, Apollo Heart Institute, Visakhapatnam, Andhra Pradesh, India
† Department of Cardiac Anaesthesia, Apollo Heart Institute, Visakhapatnam, Andhra Pradesh, India
Address correspondence to C. Sai Krishna, Department of Cardiothoracic and Vascular Surgery, Apollo Heart Institute, Waltair Main Road, Visakhapatnam 530002, Andhra Pradesh, India. E-mail: csaikrishna@yahoo.com
Roller pumps are widely used for cardiopulmonary bypass in developing nations by virtue of proven safety during several years of institutional use and cost effectiveness. However, careful adjustment of roller occlusion is needed because they are known to cause spallation, tubing wear, and the occasional incident of rupture of tubing in the extracorporeal circuit. Rupture of polyvinylchloride tubing in the pump raceway during repair of a ventricular septal defect in a 4-year-old child is discussed. The event was managed by exclusion and replacement of the defective tubing during a short period of arrest. Use of an inappropriate boot pump and failure to detect its inclusion in the bypass circuit was a significant departure from protocol. However, because occlusion settings and duration of perfusion were within acceptable limits, a manufacturing flaw could also have contributed to tubing failure, and the event may or may not have been averted by the use of larger tubing. In conclusion, this incident reiterates the need for adherence to established protocol during assembly of the pump and draws attention to the fact that tubing integrity is not a guarantee and vigilance is warranted to handle its failure.
Key words: cardiopulmonary bypass complications / cardiopulmonary bypass equipment / circulatory arrest / perfusion / spallation / tubing rupture