Issue |
J Extra Corpor Technol
Volume 47, Number 2, June 2015
|
|
---|---|---|
Page(s) | 113 - 118 | |
DOI | https://doi.org/10.1051/ject/201547113 | |
Published online | 15 June 2015 |
Technique Articles
Performance Evaluation of the Plateletworks® in the Measurement of Blood Cell Counts as compared to the Beckman Coulter Unicel DXH 800
* Department of Pathology and Laboratory Medicine, College of Medicine, University of Saskatchewan, Saskatoon Health Region, Saskatchewan, Canada
† Department of Surgery, Division of Cardiothoracic Surgery, College of Medicine, University of Saskatchewan, Saskatoon Health Region, Saskatchewan, Canada
Address correspondence to: Erick McNair, CCP, PhD, FICA, Department of Pathology and Laboratory Medicine, College of Medicine, Health Science Building, Room 2D01, Royal University Hospital, 107 Wiggins Road, University of Saskatchewan, Saskatoon, SK, S7N 5E5, Canada. E-mail: erick.mcnair@usask.ca
Received:
16
December
2014
Accepted:
1
July
2015
Prior to undergoing cardiac surgery many patients may have impaired platelet function due to platelet inhibition. Point of care testing (POCT) that produces quick results of platelet counts and function allow earlier clinician interpretation, diagnosis and treatment. Before being adopted for routine clinical use, a POCT device’s performance must be evaluated by standard laboratory techniques to ensure high quality results. The purpose of this study is to determine the performance of the Plateletworks® BC 3200 automated hematology analyzer by correlating its precision, accuracy and linearity for the measurement of blood counts to our hospital central laboratory analyzer (Beckman Coulter Unicel DXH 800). The study utilizes well described methods for Within-Run and Day-to-Day precision, comparison of methods (bias), and linearity. Control samples from the manufacturer were used for the precision studies, blood samples from 115 cardiac surgical subjects were used for comparison of methods and accuracy, and pre-diluted control samples from the manufacturer were used for the linearity studies. The precision of the Plateletworks® analyzer was acceptable. The overall coefficient of variation (CV) for the measured parameters at all levels of control for Within-Run precision was acceptable ranging from 0.65–6.4%. Likewise, the CV for the measured parameters at all levels of control for Day-to-Day precision was acceptable ranging from 1.45% to 6.7%. The correlation and accuracy between the two analyzers for the evaluated parameters (platelets, red blood cells, white blood cells, and hemoglobin) was acceptable. The linearity for the measured parameters was also acceptable with a range between 98–100%. The performance of the Plateletworks® analyzer was acceptable for providing blood cell counts as compared to our central hospital laboratory analyzer.
Key words: performance / point-of-care testing / platelet counts / correlation / precision / accuracy / linearity
© 2015 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.