Figure 2
Download original image
(a) Test circuit diagram (blood circulation before measurement). The atmospheric release circuit of the blood circuit must be emptied. The filtrate level was set at the height of the three-way stopcock for blood circuit switching, the filtration circuit was closed. The blood circuit was cycled for more than 5 min. (b) Test circuit diagram (during colloid osmotic pressure measurement). The roller pump was stopped, the three-way stopcock was switched, and the blood circuit was switched to an open-air circuit. The clamp of the filtrate circuit was then opened to expose the circuit to the atmosphere. After waiting 3 min for the solvent migration in the blood and filtrate circuits to equilibrate, the water level difference created between the two circuits was measured.