Table 2
Simultaneous reproducibility with test liquids 1–7 in Experiment 1
Measurement method (n = 10) | Mean COP [mmHg] | SD | CV [%] | |
---|---|---|---|---|
Test liquid 1 (Hct: 20.2%) | Test method | 5.77 | 3.50 × 10−2 | 0.6 |
Colloid osmometer | 5.84 | 4.90 × 10−2 | 0.8 | |
Test liquid 2 (Hct: 22.6%) | Test method | 7.09 | 1.28 × 10−1 | 1.8 |
Colloid osmometer | 7.08 | 1.10 × 10−1 | 1.5 | |
Test liquid 3 (Hct: 23.8%) | Test method | 8.23 | 1.49 × 10−1 | 1.8 |
Colloid osmometer | 8.11 | 1.00 × 10−1 | 1.3 | |
Test liquid 4 (Hct: 26.0%) | Test method | 9.77 | 8.35 × 10−2 | 0.9 |
Colloid osmometer | 9.6 | 7.70 × 10−2 | 0.8 | |
Test liquid 5 (Hct: 28.0%) | Test method | 11.65 | 9.45 × 10−2 | 0.8 |
Colloid osmometer | 11.48 | 1.10 × 10−1 | 0.9 | |
Test liquid 6 (Hct: 31.7%) | Test method | 15.51 | 4.07 × 10−1 | 2.6 |
Colloid osmometer | 14.81 | 1.80 × 10−1 | 1.2 | |
Test liquid 7 (Hct: 34.8%) | Test method | 19.69 | 3.12 × 10−1 | 1.6 |
Colloid osmometer | 18.34 | 6.60 × 10−2 | 0.4 |
Measurements were performed 10 times for each test liquid and each measurement method, and CVs (%) were calculated. For the COP measurements using the Test Method, all CVs were in the range of 1.44 ± 0.66%. For measurements obtained using the colloid osmometer, all CVs were in the range of 0.99% ± 0.34%. Abbreviations: COP, colloid osmometer; SD, standard deviation; CV, coefficient of variation; Hct, hematocrit value.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.