Table 3
Simultaneous reproducibility with test liquids 8–10 in Experiment 2.
Measurement method (n = 10) | Mean COP [mmHg] | SD | CV [%] | |
---|---|---|---|---|
Test liquid 8 (Hct: 26.2%) | Test method | 14.34 | 3.29 × 10−1 | 2.3 |
Colloid osmometer | 14.02 | 1.33 × 10−1 | 0.9 | |
Test liquid 9 (Hct: 25.9%) | Test method | 26.36 | 2.75 × 10−1 | 1.0 |
Colloid osmometer | 26.46 | 2.24 × 10−1 | 0.8 | |
Test liquid 10 (Hct: 25.8%) | Test method | 33.82 | 5.68 ×10−1 | 1.7 |
Colloid osmometer | 33.83 | 3.10 ×10−1 | 0.9 |
Measurements were performed 10 times for each test liquid and each measurement method, and CVs (%) were calculated. With COP measurement by the Test Method, all CVs were in the range of 1.67 ± 0.51%. With measurement using the colloid osmometer, all CVs were in the range of 0.90% ± 0.04%. Abbreviations: COP, colloid osmometer; SD, standard deviation; CV, coefficient of variation; Hct, hematocrit value.