Open Access
Table 2
Final list with 37 elements that were assessed as mean ≥ 4 and SD ≤ 1.02.
Items version 2 | Mean | SD |
---|---|---|
Patient | ||
01 Identification/Weight/Height | 4.4 | 0.54 |
02 Pathology | 4.4 | 0.54 |
03 Surgical procedure | 4.4 | 0.89 |
04 Calculations (heparin/flows) | 4.8 | 0.44 |
05 Blood products checking | 4.8 | 0.44 |
Equipment | ||
06 Equipment clean | 4.6 | 0.54 |
07 Power cables | 4.6 | 0.89 |
08 Heater-cooler | 4.8 | 0.44 |
CPB machine | ||
09 Initial test (Rollers/Water lines/Heat exchanger) | 4.8 | 0.44 |
10 Hand crank | 4.8 | 0.44 |
11 Level sensor | 4.0 | 0.63 |
12 Bubble sensor | 4.2 | 0.44 |
13 Battery | 4.6 | 0.89 |
Centrifugal pump | ||
14 Console/Driver/Battery | 4.4 | 0.89 |
15 Flow sensor | 4.8 | 0.44 |
16 Manual hand crank | 4.8 | 0.44 |
17 Operational test | 4.4 | 0.54 |
Gas supply | ||
18 O2/air regulation valve | 4.6 | 0.54 |
19 Lines: O2/Air (Connected) | 4.6 | 0.54 |
20 Blender/Flowmeter/FIO2 | 4.6 | 0.54 |
Materials | ||
21 Pharmacy kit | 4.6 | 0.54 |
22 Oxygenator and components | 4.4 | 0.89 |
23 Reynald clamps for tubes | 4.4 | 0.89 |
24 Ice | 4.6 | 0.89 |
Circuit preparation | ||
25 Safely connecting the tubes | 4.8 | 0.44 |
26 Perfusate | 5.0 | 0.0 |
27 Calibration of the arterial roller and aspirators | 4.4 | 0.54 |
28 Air removal from the system | 4.6 | 0.54 |
29 No leakage in the circuit | 4.4 | 0.54 |
Cardioplegia | ||
30 Test in the cardioplegia module | 4.2 | 0.83 |
31 Circuit assembly/checking | 4.4 | 0.54 |
32 Cardioplegia solution | 4.2 | 0.97 |
Vacuum system | ||
33 Vacuum regulator | 4.6 | 0.89 |
34 Relief valve | 5.0 | 0.0 |
Before Starting CPB | ||
35 Heparin dosage | 4.4 | 0.54 |
36 Anticoagulation/ACT | 4.6 | 0.54 |
37 Arterial line test | 5.0 | 0.0 |
*CPB – cardiopulmonary bypass; ACT – activated coagulation time.