Clinical Evaluation of the Sorin Synthesis™ Oxygenator With Integrated Arterial Filter
Address correspondence to: Gerard J Myers, RRT, CCP, Cardiovascular Perfusion Services, Cardiac OR, NHI Campus, QEII Health Sciences Center, PO Box 9000, Halifax, Nova Scotia Canada. B3K 6A3. E-mail: edit@ns.sympatico.ca
The use of arterial line filters has long been a standard of practice in the field of cardiopulmonary bypass. Sorin Biomedica has designed an adult hollow-fiber oxygenator that not only incorporates their Mimesys® biomimicry coating technology but also has a 40-micron arterial filter as an integrated component of this unique membrane oxygenator. We did a prospective, randomized clinical trial of 54 Synthesis® coated oxygenators and compared them with 54 uncoated Monolyth Pro® oxygenators, the latter of which incorporated an external arterial line filter with a standard bypass loop There were few statistically significant differences found between the Synthesis® group and the Monolyth group with regard to pressure differentials, hemodynamic resistance, and platelet drop. The Synthesis® oxygenator did require less priming volume, but the amount was not significant. Platelet counts with the Phosphorylcholine coated Synthesis oxygenators, using crystalloid perfusates, was similar to our previously published data on platelet protection and Albumin perfusates. We conclude that the Sorin Synthesis® oxygenator appears to have better flow characteristics than the Monolyth oxygenator, with the potential for lower priming volumes. The most clinically significant benefit comes from the elimination of the arterial filter bypass loop and the avoidance of inverting the arterial filter during priming.
Key words: synthesis / phosphorylcholine / coated / oxygenators / arterial filter / bypass loop / platelets / pressure excursion / adult / cardiopulmonary
© 2005 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.