Is the Air Handling Capability of the Quadrox D Pump Dependent within an ECMO Circuit? An In Vitro Study
Address correspondence to: Martin C. Gill, MSc, Trainee Perfusionist, Heart Centre for Children, Children’s Hospital at Westmead, Locked Bag 4001, Westmead, Sydney, New South Wales, 2145 Australia. E-mail: marting3@chw.edu.au
Received:
15
December
2009
Accepted:
19
July
2010
The occurrence of gaseous microemboli (GME) within the extracorporeal membrane oxygenation circuit is largely overlooked, as are methods to ameliorate this occurrence. We aimed to determine if the air handling capability of the Quadrox D oxygenator was dependent upon whether it was used in conjunction with a centrifugal or roller pump; and if application of a Pall air eliminating filter (AEF) would prevent circuit air introduction from intravenous infusions. Using a blood primed circuit 1 mL of air was infused pre pump. GME were quantified post pump and post oxygenator using the EDAC® Quantifier. Trials were conducted at 1 and 2 L/min flow. To prevent GME recirculation a Capiox SX18 was used in circuit with negative pressure applied to its oxygenator; an EDAC® cuvette distal to this device quantified GME recirculation. Following air infusion, 3–5 minute data recordings were carried out for each trial. Separate trials were carried out for centrifugal and roller pumps, and for each flow rate. The process was then repeated following the application of the AEF to the air infusion line. More GME were detected post Quadrox D when the centrifugal pump was used in comparison to the roller pump at 1 L/min (p ≤ .05), and 2 L/min (p = .05). A greater volume of air was detected post Quadrox D when used in conjunction with the centrifugal device at 1 L/min (p ≤ .05), and 2 L/min (p ≤ .05). Application of the AEF resulted in zero GME detected at any circuit location. The results of this study confirm that a greater total count and volume of GME are detected distal to the Quadrox D when used in conjunction with a Rotaflow centrifugal pump. Application of a Pall AEF to infusion and drug lines can prevent air introduction from this source.
Key words: gaseous microemboli / oxygenator / centrifugal pump / roller pump / extracorporeal membrane oxygenation / EDAC®
© 2010 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.