Issue |
J Extra Corpor Technol
Volume 44, Number 3, September 2012
|
|
---|---|---|
Page(s) | 126 - 133 | |
DOI | https://doi.org/10.1051/ject/201244126 | |
Published online | 15 September 2012 |
Original Articles
Endogenous Gas Formation—An In Vitro Study with Relevance to Gas Microemboli during Cardiopulmonary Bypass
Address correspondence to: Lena Lindholm, PhD, ECCP, Heart Centre, Cardiothoracic Surgery, Department of Surgery and Perioperative Sciences, University of Umeå, S-901 85 Umeå, Sweden. E-mail: lena.e.lindholm@vll.se
Received:
14
December
2011
Accepted:
15
June
2012
Gas embolismis an identified problemduring cardiopulmonary bypass (CPB). Our aim was to analyze the potential influence from gas solubility based on simple physical laws, here called endogenous gas embolism. Gas solubility decreases at higher temperature and gas bubbles are presumably formed at CPB warming. An experimental model to measure gas release was designed. Medium (water or blood retrieved from mediastinal drains, 14.6 mL) was incubated and equilibrated with gas (air, 100% oxygen, or 5% carbon dioxide in air) at low temperature (10°C or 23°C). At warming to 37°C, gas release was digitally measured. Also, the effect of fluid motion was evaluated. At warming, the medium became oversaturated with dissolved gas. When fluid motion was applied, gas was released to form bubbles. This was exemplified by a gas release of .45% (.31/.54, medians and quartile range, volume percent, p = .007) and 1.26% (1.14/1.33, p = .003) when blood was warmed from 23°C or 10°C to 37°C, respectively (carbon dioxide 5% in air). Consistent findings were seen for water and with the other types of gas exposure. The theory of endogenous gas embolization was confirmed with gas being released at warming. The endogenous gas formation demonstrated a dynamic pattern with oversaturation and with rapid gas released at fluid motion. The gas release at warming was substantial, in particular when the results were extrapolated to full-scale CPB conditions. The interference from endogenous gas formation should be considered in parallel to external sources of gas microemboli.
Key words: cardiopulmonary bypass / gas embolization / microemboli / gas solubility / temperature
© 2012 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.