Brief Communication
One-Way Valve Malfunction in an Extracorporeal Membrane Oxygenation Priming Circuit
Address correspondence to: W. Cory Ellis, CCP, Perfusion, Children’s Hospital Colorado, 13123 East 16th Avenue, Aurora, CO 80045. E-mail: william.ellis@childrenscolorado.org
Received:
18
September
2013
Accepted:
18
February
2014
Developing technologies have changed both the components and the management style when extracorporeal membrane oxygenation (ECMO) is used to support critically ill cardiac and respiratory patients. The Cardiohelp system™ is a small, portable extracorporeal system just recently available within the United States. Manufacturing standards and quality processes have made mechanical failure and malfunction of extracorporeal components less common; however, there is still potential for mechanical failure or component malfunction before or during extracorporeal support. This case review describes the malfunction of a Retroguard™ unidirectional flow valve integrated into the priming setup of a Cardiohelp system™ during the priming process.
Key words: ECMO / Cardiohelp / Retroguard / one-way valve / mechanical failure
© 2014 AMSECT
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.