Designing an experimental method for assessing biocompatibility of circuit coatings using biomarkers for platelet activation during cardiopulmonary bypass
1
Arizona College of Osteopathic Medicine, Midwestern University, 19555 N. 59th Avenue, Glendale, AZ 85308, USA
2
Office of Research & Sponsored Program, Midwestern University, 19555 N. 59th Avenue, Glendale, AZ 85308, USA
3
College of Veterinary Medicine, Midwestern University, 19555 N. 59th Avenue, Glendale, AZ 85308, USA
4
College of Graduate Studies, Midwestern University, 19555 N. 59th Avenue, Glendale, AZ 85308, USA
5
College of Health Sciences, Midwestern University, 19555 N. 59th Avenue, Glendale, AZ 85308, USA
6
College of Medicine Phoenix, University of Arizona, 19555 N. 59th Avenue, Glendale, AZ 85308, USA
7
Faculty of Medicine, University of British Columbia, 2176 Health Sciences Mall, Vancouver, BC V6T 2A1, Canada
* Corresponding authors: ndarba@midwestern.edu; mesfan@midwestern.edu
Received:
2
February
2023
Accepted:
10
February
2024
Introduction: Cardiopulmonary bypass is an essential component of cardiothoracic surgeries. However, significant complications such as systemic inflammatory response syndrome (SIRS) resulting from cardiopulmonary bypass (CPB) are a common occurrence due to contact between circulating blood and foreign surfaces that leads to platelet activation. It is suggested that different available CPB circuit coatings can potentially reduce platelet activation. However, there have been no published evidence-based reports confirming these claims. In addition, there is no well-established protocol for studying platelet activation biomarkers during CPB in vitro in a laboratory setting. Methods: CPB was simulated in the laboratory using bovine blood in two different types of coated CPB circuits: Trillium® Biosurface by Medtronic, and XcoatingTM Surface by Terumo. Fresh bovine blood samples were collected and circulated through the CPB circuit following the standard protocol used in the operation rooms. Blood samples were then collected at 5 min, 30 min, and 55 min during the circulation. Blood plasmas were separated and subjected to enzyme-linked immunosorbent assay to measure most established platelet activation markers P-selectin, Platelet Factor 4 (PF4), Glycoprotein IIb/IIIa (GPIIb/IIIa), and β-thromboglobulin (β-TG) at different time points. Results: The biomarker values at 30 min and 55 min were compared to the base values at 5 min for each type of CPB circuit. The results of the means from all measured biomarkers showed data measurements that indicated no significant variability within each coating. All collected data points fell within ±2 SD of the means, which was considered acceptable variations across technical replicates. Conclusion: In this study, we were able to establish an in vitro protocol in the laboratory setting that is precise and reliable with minimum intra-variability. This established protocol will allow for future studies in which different coated CPB circuits can be compared for their effectiveness in blocking platelet activation during the CPB.
Key words: Cardiopulmonary bypass / Platelets / Biomarkers / Platelet activation / ELISA / Circuit coating
© The Author(s), published by EDP Sciences, 2024
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Introduction
Cardiopulmonary bypass (CPB) provides a bloodless, motionless field for the surgeon to operate. While this has facilitated many advances in the field, it does pose significant complications, such as platelet activation and dysfunction, as well as abnormal bleeding [1]. Serious manifestations of these complications can lead to systemic inflammatory response syndrome (SIRS) and sepsis, leading to organ failure [2]. Most of these complications can be traced back to platelet activation in contact with a foreign body, such as the coating of circuits used in CPB. During the bypass, platelets bind to these artificial surfaces, leading to the release of inflammatory cytokines and activation of coagulation-fibrinolysis systems [3]. The surface markers of platelets such as platelet factor 4 (PF4), β-thromboglobulin (β-TG), glycoprotein IIb/IIIa (GPIIb/IIIa), and soluble P-selectin are released during the coagulation cascade and are considered as reliable markers for platelet activation during CPB [4].
Recent advances in bypass technologies have created biocompatible surfaces to potentially decrease platelet activation during CPB, therefore, reducing the number of complications after surgery. Heparin-coated circuits are widely used because they seem not to cause excessive platelet activation as compared to non-coated circuits [5]. Another commonly used circuit poly-2-methoxyethyl acrylate (PMEA) has a protective water layer that allows proteins to retain their conformation and reduce adherence to foreign surfaces [6].
While this has been the focus of research and advancement in cardiothoracic surgery, there is minimal scientific and data-driven evidence that establishes the biocompatibility of these circuits or their potential benefits in reducing platelet activation during CPB; hence, there has yet to be a reliable and standardized laboratory protocol that compares these biocompatible circuits in vitro with respect to their effects on platelet function and activation [7].
Much of our evidence of biocompatibility comes from platelet activation studies done during clinical studies with patients undergoing surgery with limited sample size and multiple independent variables [8–10]. In this study, we aimed to 1) establish a well-controlled and standardized in vitro protocol to measure surface markers of platelet activation at different time points during the bypass in the laboratory setting using commercially available Enzyme-Linked Immunosorbent Assay (ELISA) kits, and 2) determine the intra-coating variability across technical replicates for two types of coated circuit that are commonly used in CPB to establish the required technical replicate within each independent biological replicate.
Materials and methods
Bovine blood collection
Bovine blood was obtained via venipuncture from a local slaughterhouse and treated with 30,000 IU of heparin in the collection bucket to reduce clotting during sample transportation. Eight to nine (8–9) liters of bovine blood were used for one run of CPB. An activated clotting time was accepted and maintained above 480 s after heparin administration [11]. The circuit was not connected to a living system; hence no extra heparin was added during the experimental runs of CPB. Experiments with CPB were started within 1–2 h and completed within 8–9 h post-blood collection.
CPB circuit setup
One bovine blood donation was used for five technical replicates of experimental CPB runs using Trillium ® Biosurface (541T, Medtronic Inc., Minneapolis, MN) with a heparinized surface. Another bovine blood donation was used for five technical replicates of experimental CPB runs using XcoatingTM (3CX*FX25RWC, Terumo Cardiovascular, Ann Arbor, MI), which is a PMEA-coated circuit [12]. The laboratory simulated circuit included a pump, arteriovenous (AV) loops, tubing, reservoirs, oxygenators, arterial line filters, a heat exchanger (16-02-85, SORIN Group Deutschland GmbH, Müchen, Germany), and a roller pump (Stockert S3 Roller Pump with S3 Console 10-60-00, SORIN Group Deutschland GmbH, Müchen, Germany). All parts of the circuit were replaced before each technical replicate trial (Fig. 1).
Figure 1 Schematic of the laboratory-based (in vitro) Cardiopulmonary Bypass Circuit. |
During the experimental runs, bovine blood was mixed with crystalloid (Normosol-R) to prime and de-air the CPB circuit. The amount of crystalloid used was calculated to maintain a hematocrit of 23–27%. Blood was circulated at a rate of 4 L/min and was mixed with crystalloid for the first 5 min of the run. Blood samples were collected at baseline (5 min after initiation of CPB at 37 °C), 30 min after initiation of CPB at 32 °C, and 55 min after initiation of CPB at 37 °C. The circuit run was completed in 60 min. At each desired time point, 1 mL of blood samples were collected using K2 EDTA-coated Vacutainer tubes (catalog number 367863; Becton Dickinson, San Diego, CA). Collected blood samples were centrifuged at 2500 RPM (Microfuge Lite 367121, Beckman Coulter, Inc., Palo Alto, CA) for 15 min. Separated plasma samples were aliquoted (500 μL) into small CryTubes (Thermo Scientific Nalgene System, Rochester, NY) and stored at −80 °C for ELISA assay.
Measurements of platelet count
The whole bovine blood sample was collected in a sterile vacutainer tube containing 15% EDTA K3 solution (Covidien, Cardinal Health, 8881311446) and stored at 4 °C. Platelet counts were assessed using Siemens Advia 2120i (Siemens Medical Solutions, Inc., Malvern, PA) by a single optical cytometer after appropriate dilution of blood samples with ADVIA series RBC/PLT reagent. Platelets were counted from the signals of the common detector, and coincidence correction was made so that accurate counts were made over a wide range of cell types.
Assessments of platelet activation biomarkers
Markers for platelet activation were measured in duplicates by ELISA assay according to instructions provided by the manufacturer (MyBioSource Inc., San Diego, CA) and at a 1:2 dilution. Collected plasma samples were used to measure four biomarkers of platelet activation: PF4 (MBS741915), β-TG (MBSO89872), P-selectin (MBS734634), and GPIIb/IIIa (MBS739298) at three different collection time points (5 min, 30 min, and 55 min). Detection ranges and sensitivity for each kit are as listed: PF4 (25–500 ng/mL and 1.0 ng/mL), β-TG (0.156–10 ng/mL and 0.05 ng/mL), P-selectin (5–100 ng/mL and 1.0 ng/mL), and GPIIb/IIIa (1–25 ng/mL and 0.1 ng/mL). The values for standards and experimental samples were then analyzed using the ELISA analysis software (BioTek ELISA Version 2.06, BioTek, Winooski, VT).
Statistical analysis
The values obtained from the ELISA assays were averaged across the five runs within each of the two coatings for each time point at 5 min, 30 min, and 55 min. The analysis methods used were descriptive statistics to quantify potential changes in biomarker values as determined by the ELISA assays. A standard deviation (SD) from the means was calculated and the individual values of each run were compared to the mean to see if they were within ±2 SD of the mean. This method was used to determine whether any outlier values were found. Error bars were calculated as standard errors, and the results were represented as bar graphs. No statistical inference tests between two different coatings were used given the limited sample size and low statistical power. The descriptive statistics from this pilot study will be used to inform and power future study designs with adequate biological replicates.
Results
Quantification of platelet counts
To establish a reliable in vitro protocol to measure platelet count post bypass using different coatings, we measured the range for inter-variation between bypass technical replicates for Trillium® Biosurface and XcoatingTM circuits, by running the same bovine blood samples five times through each type of coated circuits (Fig. 2). A new package of circuit was used for each technical replicate run (five unused circuit packages for each type of coating) and all five replicates were run during the same experimental day. Our data showed no significant increase or drop in platelet count at 5 min, 30 min, and 55 min for both coating types, with values at each time point falling within ±2 SD of the mean for Trillium® Biosurface (Fig. 2A) and XcoatingTM (Fig. 2B).
Figure 2 Quantification of Platelet counts in circulated bovine blood. (A) Bar graph presentation of platelet counts at different time points of bovine blood circulation using Trillium® Biosurface circuit. Data presents the average of five technical replicates using the same blood samples at 5 min, 30 min, and 55 min with no significant changes between time points. (B) Bar graph presentation of platelet counts at different time points of bovine blood circulation using XcoatingTM circuit. Data presents the average of five technical replicates using the same blood samples at 5 min, 30 min, and 55 min with no significant changes between time points. (Mean ± SD, P < 0.05, five technical replicates). |
Measurements of biomarkers of platelet activation in Trillium® Biosurface circuit
ELISA assay was used to measure the levels of platelet activation markers (soluble P-selectin, PF4, GPIIb/IIIa, and β-TG) in circulated bovine blood at 5-, 30-, and 55-minutes during circulation using Trillium® Biosurface coated tubes. The values for biomarkers at 5 min were arbitrarily set as the baseline measurement for each biomarker within the same circuit type. Measurements at 30 min and 55 min were then compared to measured values at 5 min to assess potential time-dependent changes in biomarkers at each collection time point. All the biomarker concentrations were normalized over the platelet counts for that time point. The results indicate that in the Trillium® Biosurface circuit, the values for platelet activation biomarkers fall within ±2 SD of the mean for soluble P-selectin, GPIIb/IIIa, PF4, and β-TG (Fig. 3). The normalized values for soluble P-selectin dropped at 30 min and 55 min (Fig. 3A), with GPIIb/IIIa levels showing a significant drop only at 55 min (Fig. 3B) as compared to basal levels recorded at 5 min. On the other hand, the values for PF4 and β-TG did not show any significant changes with time (Figs. 3C and 3D). The data indicates that the luminal surface of Trillium® Biosurface circuits has no effects on the secretion levels of these two biomarkers in circulated blood. We also ran statistical analyses using the recorded raw values for all four biomarkers secretion and translocation levels without normalization over the platelet counts in bypassed bovine blood at 5–, 30–, and 55–minutes in Trillium® Biosurface circuit (Supplementary Fig. S1).
Figure 3 Measurements of biomarkers of platelet activation in Trillium® Biosurface circuit. (A) Soluble P-selectin levels reduced at 30 min and 55 min, with values at each time point falling within ±2 SD of the mean. All measurements are presented as normalized values over the platelet count for each time points. (B) The normalized values for GPIIb/IIIa levels in circulated bovine blood show a significant decrease at 55 min, with all measured values falling within ±2 SD of the mean. (C) Normalized values for PF4 and (D) β-TG did not show any significant changes at 5 min, 30 min, and 55 min, but all values fell within ±2 SD of the mean, indicating minimal intra-variability across technical replicates. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin. |
Measurements of biomarkers of platelet activation in XcoatingTM Circuit
The same blood collection methods and time points were used to evaluate the secretion and translocation levels of four platelet activation biomarkers using the XcoatingTM bypass circuit. Our results show that normalized values (over platelet counts) for soluble P-selectin and PF4 translocation and release transiently but not significantly increased by 30 min, and returned to basal levels, with no statistically significant changes observed at 55 min (Figs. 4A and 4B), suggesting that luminal surface of XcoatingTM tubes do not impact the level of soluble P-selectin and PF4 release in circulated blood. On the other hand, normalized values for GPIIb/IIIa showed a significant drop at 55 min in the XcoatingTM circuit (Fig. 4C), with no significant changes observed for β-TG levels in the circulated blood at any time points (Fig. 4D). In addition, statistical analyses were performed using raw values for all four biomarkers secretion and platelet surface levels without normalization over the platelet counts in bypassed bovine blood at 5-, 30-, and 55-minutes in XcoatingTM bypass circuit (Supplementary Fig. S2). A summarized version of all collected values for platelet activation markers in circulated bovine blood is also presented in Table 1.
Figure 4 Measurements of biomarkers of platelet activation in XcoatingTM circuit. (A) No significant changes were detected in soluble P-selectin levels in collected bovine blood at 5 min, 30 min, and 55 min using the XcoatingTM circuit, with values at each time point falling within ±2 SD of the mean. All measurements are presented as normalized values over the platelet count for each time point. (B) The normalized values for PF4 did not change over time with all values at each timepoint falling within ±2 SD of the mean. (C) Normalized values for GPIIb/IIIa and (D) β-TG did not show any significant changes at 5 min, 30 min, and 55 min, but all values fell within ±2 SD of the mean, indicating minimal intra-variability across technical replicates. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin. |
A collective list of values for four biomarkers of platelet activation using Trillium® Biosurface and XcoatingTM circuits. GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin
Discussion
A cardiopulmonary bypass is a machine that allows for a surgical field suitable to perform open heart surgery, however, it is not devoid of complications [1]. CPB-induced SIRS affects up to 10% of patients and can lead to hypoperfusion, embolization, multiple organ failure, and death [13]. Several proinflammatory pathways work synergistically to cause tissue destruction, and abnormal bleeding [13]. Platelets are known to bind damaged blood vessels and cause coagulation [4]. Therefore, in cases of serious complications like CPB-induced SIRS, dysfunctional platelets are the likely reason for organ damage and excessive bleeding. Modern technology has introduced heparin-coated and PMEA-coated circuits to reduce platelet activation and control some of the complications associated with CPB. However, there is no established and well-controlled in vitro protocol that allows for the accurate assessment of the efficiency of commercially available bio-coated bypass circuits in blocking the CPB-mediated platelet activation in a surgical setting.
The main objective of this study was to establish a standardized method to study platelet activation in vitro for future comparative studies, and to further determine if there are any significant inter-variations across the circuit batches that would potentially impact the accuracy of not only the biomarker measurements across technical and biological replicates, but also the reliability of our statistical analyses. Moreover, the data collected in the current pilot study will allow for appropriate study sample size calculation that provides enough power for more reliable and reproducible comparative future studies across different types of biocoated CPB circuits.
In this study, we measured platelet counts and the secretion and translocation levels of known and well-established biomarkers of platelet activation including PF4, β-TG, GPIIb/IIIa, and soluble P-selectin at different time points of CBP. Together, these biomarkers provide a comprehensive profile of platelet activation, aggregation, adhesion, and subsequent coagulation cascade that is triggered by the contact of platelets with foreign surfaces such as the lumen of the CPB circuits.
PF4 is a cytokine released from alpha granules of platelets during activation and blood coagulation, and its increased level was found to be consistent with the incidence of cardiac complications during bypass surgeries [3]. In addition, PF4 was shown to be elevated in peripheral artery disease, acute ischemic stroke, essential thrombocytopenia, and other clinical conditions that were associated with platelet activation and is thus considered a relatively more sensitive marker of platelet activation during pathological events [14]. XcoatingTM showed a decrease in PF4 total amount throughout the trial, however, when normalized with platelet count, there was an increase then decrease in PF4. Although there was an increase in PF4 normalized concentration, it remained unchanged at the end of the trial. Trillium® Biosurface showed a constant PF4 concentration indicating that no further activation occurred to increase PF4.
In addition to PF4, β-TG is another cytokine released early following platelet adhesion and is commonly used as a reliable marker of platelet activation in vitro [14]. After the initial increase in β-TG concentration, Trillium® Biosurface and XcoatingTM did not have significant effects throughout the experiments. Another platelet marker GPIIb/IIIa belongs to the large family of integrin complexes and plays a critical role in platelet aggregation and increased binding to plasma fibrinogen and endothelial von Willebrand factor, therefore, facilitating thrombin generation and blocking hemorrhage [15]. Furthermore, GPIIb/IIIa surface translocation increases upon platelet activation making it a valuable biomarker in quantifying platelet activation and aggregation [16]. Both circuit coatings prevented further GPIIb/IIIa activation indicated by a decrease in concentration during the trials. Another platelet activation marker P-selectin is stored in alpha granules of platelets as well as in the Weibel-bodies of endothelial cells that cover the lumen of the blood vessels. Upon activation, P-selectin is translocated to the surface of endothelium and released into the blood circulation and plays an important role during the initial adhesion and rolling of platelets in the site of injury [17, 18]. In addition, activated platelets result in an increase in soluble P-selectin translocation and release allowing for quantification of platelet activation [19]. The inhibition of P-selectin release was displayed by a decrease in concentration over the course of the trial in Trillium® Biosurface. When normalized over the platelet counts, soluble P-selectin levels did not show any significant changes throughout the experiment using XcoatingTM circuits.
This was a longitudinal study that measured the platelet activation at different time points during a circuit run for two different types of CPB circuit coatings. The baseline measurement was arbitrarily set at 5 min, so there would be adequate time for the crystalloid to mix with the bovine blood within the circuit. Measurements done at 30 min and 55 min were compared to 5 min and shown to both fall within ±2 standard deviations of the mean at each time point for each type of coated circuit. ELISA assays were then utilized to measure platelet activation markers due to their high sensitivity and specificity. While these assays are not intended to be used in a clinical setting for patients undergoing CPB, they can provide insight into the activity of platelets within the CPB coating during cardiac surgery.
It is noteworthy, that although this study has shown minimum inter-variations within each type of coating, it cannot provide conclusive comparative data of any potential differences between the coating types. The present study was designed to establish a standardized protocol in a laboratory setting that would allow us to plan future experiments to compare different types of bio-coated circuits in the market and address a very critical question: which type of commercially available CPB circuit would provide better protection against platelet activation during human bypass surgeries? In addition, we were not able to have a positive control group for this study, as uncoated reservoirs or tubing are not available from the same vendor for clinical use. The use of custom-made uncoated CPB circuit would have added confounding variables to the study. Our established protocol remains accurate shown by the similarities in biomarker activation in two different circuit coatings and the precision, consistency, and duplicability of the ELISA results.
In this study, we were able to establish a laboratory protocol using bovine blood that minimizes the inter-variations within each type of circuit through different time points. The results presented in this report confirm that the method is precise, reliable, and reproducible given no inter-variation within each coating at different time points. The preliminary data collected in this study will be used for power analysis and sample size calculations to determine the proper sample size (numbers for technical and biological replicates) for future comparative studies.
Funding
This study was supported by Midwestern University Graduate Fund (K.N), college of Health Sciences Faculty Intermural Grant (N.D.), and Arizona College of Osteopathic Medicine KAS Summer Research Fellowship Award (M.S.).
Conflict of interest
The authors declare no conflict of interest.
Data availability statement
The raw original data used in this study is available by the authors upon request.
Authors contribution statement
MS contributed to data generation, data analysis, and manuscript writing. MR had equal contribution to data generation, analysis, and manuscript writing. CB helped with experimental design, data analysis, and data presentation. WL conducted the ELISA test and helped with data generation and analysis. KN helped with experimental protocol and data collection. TR helped with data interpretation and manuscript writing. ME and ND are senior investigators involved in experimental design, data analysis and review, manuscript writing and editing.
Ethics approval
The study does not involve animal or human subjects. The biosafety protocol was approved by Midwestern University Office of Research & Sponsored Program.
Supplementary material
Supplementary Figure S1. Raw values for platelet activation biomarkers using Trillium® Biosurface bypass. Bar graphs presenting raw values of platelet activation markers in collected bovine blood at 5-, 30-, and 55-minutes using Trillium® Biosurface bypass circuit without normalization over the platelet counts. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin |
Supplementary Figure S2. Raw values for platelet activation biomarkers using XcoatingTM bypass. Bargraphs presenting raw values of platelet activation markers in collected bovine blood at 5-, 30-, and 55-minutes using XcoatingTM bypass circuit without normalization over the platelet counts. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin |
References
- Bartoszko J, Karkouti K. Managing the coagulopathy associated with cardiopulmonary bypass. J Thromb Haemost 2021;19(3):617–632. [CrossRef] [PubMed] [Google Scholar]
- Maisat W, Yuki K. Narrative review of systemic inflammatory response mechanisms in cardiac surgery and immunomodulatory role of anesthetic agents. Ann Card Anaesth 2023;26(2):133–142. [CrossRef] [PubMed] [Google Scholar]
- Fukui T, Nishida H, Takanashi S. Biocompatibility of cardiopulmonary bypass circuit with new polymer Senko E-Ternal Coating. Perfusion 2015;30(7):572–9. [CrossRef] [PubMed] [Google Scholar]
- Yun SH, Sim EH, Goh RY, Park JI, Han JY. Platelet activation: the mechanisms and potential biomarkers. Biomed Res Int 2016;2016:9060143. [PubMed] [Google Scholar]
- Hanedan MO, Yürük MA, Arslan AK, Kılıç A, Sayar U, Mataracı İ. Heparin-coated vs. non-coated cardiopulmonary bypass circuits: comparing immediate results with different target activated clotting time. Braz. J Cardiovasc Surg 2020;35(6):913–917. [Google Scholar]
- Hosoyama K, Ito K, Kawamoto S, et al. Poly-2-methoxyethylacrylate-coated cardiopulmonary bypass circuit can reduce transfusion of platelet products compared to heparin-coated circuit during aortic arch surgery. J Artif Organs 2016;19(3):233–40. [CrossRef] [PubMed] [Google Scholar]
- Lu Q, Hofferbert BV, Koo G, Malinauskas RA. In vitro shear stress-induced platelet activation: sensitivity of human and bovine blood. Artif Organs 2013;37(10):894–903. [CrossRef] [PubMed] [Google Scholar]
- Lax M, Pesonen E, Hiippala S, Schramko A, Lassila R, Raivio P. Heparin dose and point-of-care measurements of hemostasis in cardiac surgery-results of a randomized controlled trial. J Cardiothorac Vasc Anesth 2020;34(9):2362–2368. [CrossRef] [PubMed] [Google Scholar]
- Vukicevic P, Klisic A, Neskovic V, et al. New markers of platelet activation and reactivity and oxidative stress parameters in patients undergoing coronary artery bypass grafting. Oxid Med Cell Longev 2021;2021:8915253. [PubMed] [Google Scholar]
- Epstein D, Vishnepolsky A, Bolotin G, Atweh N, Bonstein L, Lehavi A. Effect of prolonged hypothermic cardiopulmonary bypass, heparin, and protamine on platelet: a small-group study. Thorac Cardiovasc Surg 2021;69(8):719–722. [CrossRef] [PubMed] [Google Scholar]
- Jamiolkowski MA, Patel M, Golding MD, Malinauskas RA, Lu Q. Comparison of animal and human blood for in vitro dynamic thrombogenicity testing of biomaterials. Artif Organs 2022;46(12):2400–2411. [CrossRef] [PubMed] [Google Scholar]
- Kocakulak M, Koçum C, Saber R, et al. Investigation of blood compatibility of PMEA coated extracorporeal circuits. J Bioact Compat Polym 2002;17(5):343–356. [CrossRef] [Google Scholar]
- Jenke A, Yazdanyar M, Miyahara S, et al. AdipoRon attenuates inflammation and impairment of cardiac function associated with cardiopulmonary bypass-induced systemic inflammatory response syndrome. J Am Heart Assoc 2021;10(6):e018097. [CrossRef] [PubMed] [Google Scholar]
- Nagy B Jr., Debreceni IB, Kappelmayer J. Flow cytometric investigation of classical and alternative platelet activation markers. EJIFCC 2013;23(4):124–34. [PubMed] [Google Scholar]
- French DL, Seligsohn U. Platelet glycoprotein IIb/IIIa receptors and Glanzmann’s thrombasthenia. Arterioscler Thromb Vasc Biol. 2000;20(3):607–10. [CrossRef] [PubMed] [Google Scholar]
- Chen Z, Mondal NK, Ding J, et al. Activation and shedding of platelet glycoprotein IIb/IIIa under non-physiological shear stress. Mol Cell Biochem 2015;409(1–2):93–101. [CrossRef] [PubMed] [Google Scholar]
- Ruggeri ZM, Mendolicchio GL. Adhesion mechanisms in platelet function. Circ Res 2007;100(12):1673–85. [CrossRef] [PubMed] [Google Scholar]
- Neubauer K, Zieger B. Endothelial cells and coagulation. Cell Tissue Res 2022;387(3):391–398. [CrossRef] [PubMed] [Google Scholar]
- Fox SC, May JA, Dovlatova N, et al. How does measurement of platelet P-selectin compare with other methods of measuring platelet function as a means of determining the effectiveness of antiplatelet therapy? Platelets 2019;30(3):290–295. [CrossRef] [PubMed] [Google Scholar]
Cite this article as: Sancheti M, Rentschler M, Bolch C, Li W, Necco K, Rath T, Esfandiarei M & Darban N. Designing an experimental method for assessing biocompatibility of circuit coatings using biomarkers for platelet activation during cardiopulmonary bypass. J Extra Corpor Technol 2024, 56, 37–44
All Tables
A collective list of values for four biomarkers of platelet activation using Trillium® Biosurface and XcoatingTM circuits. GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin
All Figures
Figure 1 Schematic of the laboratory-based (in vitro) Cardiopulmonary Bypass Circuit. |
|
In the text |
Figure 2 Quantification of Platelet counts in circulated bovine blood. (A) Bar graph presentation of platelet counts at different time points of bovine blood circulation using Trillium® Biosurface circuit. Data presents the average of five technical replicates using the same blood samples at 5 min, 30 min, and 55 min with no significant changes between time points. (B) Bar graph presentation of platelet counts at different time points of bovine blood circulation using XcoatingTM circuit. Data presents the average of five technical replicates using the same blood samples at 5 min, 30 min, and 55 min with no significant changes between time points. (Mean ± SD, P < 0.05, five technical replicates). |
|
In the text |
Figure 3 Measurements of biomarkers of platelet activation in Trillium® Biosurface circuit. (A) Soluble P-selectin levels reduced at 30 min and 55 min, with values at each time point falling within ±2 SD of the mean. All measurements are presented as normalized values over the platelet count for each time points. (B) The normalized values for GPIIb/IIIa levels in circulated bovine blood show a significant decrease at 55 min, with all measured values falling within ±2 SD of the mean. (C) Normalized values for PF4 and (D) β-TG did not show any significant changes at 5 min, 30 min, and 55 min, but all values fell within ±2 SD of the mean, indicating minimal intra-variability across technical replicates. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin. |
|
In the text |
Figure 4 Measurements of biomarkers of platelet activation in XcoatingTM circuit. (A) No significant changes were detected in soluble P-selectin levels in collected bovine blood at 5 min, 30 min, and 55 min using the XcoatingTM circuit, with values at each time point falling within ±2 SD of the mean. All measurements are presented as normalized values over the platelet count for each time point. (B) The normalized values for PF4 did not change over time with all values at each timepoint falling within ±2 SD of the mean. (C) Normalized values for GPIIb/IIIa and (D) β-TG did not show any significant changes at 5 min, 30 min, and 55 min, but all values fell within ±2 SD of the mean, indicating minimal intra-variability across technical replicates. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin. |
|
In the text |
Supplementary Figure S1. Raw values for platelet activation biomarkers using Trillium® Biosurface bypass. Bar graphs presenting raw values of platelet activation markers in collected bovine blood at 5-, 30-, and 55-minutes using Trillium® Biosurface bypass circuit without normalization over the platelet counts. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin |
|
In the text |
Supplementary Figure S2. Raw values for platelet activation biomarkers using XcoatingTM bypass. Bargraphs presenting raw values of platelet activation markers in collected bovine blood at 5-, 30-, and 55-minutes using XcoatingTM bypass circuit without normalization over the platelet counts. (Mean ± SD, P < 0.05, five technical replicates). GPIIb/IIIa: Glycoprotein IIb/IIIa; PF4: Platelet Factor 4; β-TG: β-Thromboglobulin |
|
In the text |
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.